Review and standardization of cell phone exposure calculations using the SAM phantom and anatomically correct head models

Review and standardization of cell phone exposure calculations using the SAM phantom and anatomically correct head models

Brian B Beard* and Wolfgang Kainz

  • * Corresponding author: Brian B Beard brian.beard@fda.hhs.gov

Author Affiliations

Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration, Rockville, Maryland, USA

For all author emails, please log on.

BioMedical Engineering OnLine 2004, 3:34 doi:10.1186/1475-925X-3-34

 

The electronic version of this article is the complete one and can be found online at: http://www.biomedical-engineering-online.com/content/3/1/34

 

Received: 21 May 2004
Accepted: 13 October 2004
Published: 13 October 2004

 

© 2004 Beard and Kainz; licensee BioMed Central Ltd.

This is an open-access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

We reviewed articles using computational RF dosimetry to compare the Specific Anthropomorphic Mannequin (SAM) to anatomically correct models of the human head. Published conclusions based on such comparisons have varied widely. We looked for reasons that might cause apparently similar comparisons to produce dissimilar results. We also looked at the information needed to adequately compare the results of computational RF dosimetry studies. We concluded studies were not comparable because of differences in definitions, models, and methodology. Therefore we propose a protocol, developed by an IEEE standards group, as an initial step in alleviating this problem. The protocol calls for a benchmark validation study comparing the SAM phantom to two anatomically correct models of the human head. It also establishes common definitions and reporting requirements that will increase the comparability of all computational RF dosimetry studies of the human head.

1. Background

Cell phone safety remains a topic of broad public concern that attracts frequent media attention. This attention is focused on two areas of scientific controversy concerning cell phone safety. The first area is that of non-thermal biological effects. The existence of these effects is an important open question, but it is not the topic of this paper. However, if these effects exist, their manifestation will certainly be related to the amount of RF energy deposited in the tissue – RF dosimetry [1]. The second area of controversy, and the topic of this paper, is that of RF dosimetry, specifically computational RF dosimetry. Simply put, this is a computer simulation that estimates the deposition of RF energy, the specific absorption rate (SAR), in the head of a user. Because live human heads can not be safely instrumented for these measurements, computational RF dosimetry provides the best estimate of SAR in actual human heads. For this same reason, compliance testing is done with phantom heads.

The phantom head that is now the world-wide standard for compliance testing is the Specific Anthropomorphic Mannequin (SAM). SAM was developed by members of IEEE Standards Coordinating Committee 34, SubCommittee 2, Working Group 1 (SCC34/SC2/WG1). This working group was created to develop recommended practices for determining SAR in the head via measurement techniques [2]. SAM has also been adopted by the European Committee for Electrical Standardization (CENELEC) [3], the International Electrotechnical Commission [4], Association of Radio Industries and Businesses [5], and Federal Communications Commission [6].

-review the complete article here-

In summary, the authors of this excellent review provide insight to cell phone exposure calculations.  They provide even more incentive for certain areas to restrict cell phone use without the need for cell phone jammer implementation to enforce non-compliance.